Cleaning examination cameras
GEIS camera, exam glasses, digital otoscope and intraoral dental camera
What this is: cleaning steps for Visionflex examination cameras.
Who it's for: clinical staff who use the cameras during consultations.
Cameras that go in the mouth or ear need a fresh single-use barrier for every patient, then a wipe of the device afterwards. The barrier does most of the infection control work. The wipe catches what the barrier misses. The manufacturer's instructions for use (IFU) are the final word on cleaning each device. Where this article and the IFU differ, follow the IFU. Your organisation's infection prevention policy applies on top of both.
Visionflex GEIS General Examination Camera HD
Non-critical for skin and wound use. Semi-critical in the mouth with the tongue depressor adapter. The camera is IP54 rated, which protects it from splashes but not from soaking.
| Part | How to clean | How often |
|---|---|---|
| Camera body | Clinell or a cloth lightly dampened with 70% IPA or ethanol. Keep liquid out of the USB connector and lens opening. | After each patient |
| Derma Hood (reusable) | Clinell or 70% IPA wipe | After each patient |
| Derma Hood Cap | Single use. Fit a new cap before each skin examination. | Each patient |
| Tongue depressor adapter | Clinell or 70% IPA wipe. The wooden depressor is single use. | Each patient |
| Reference stick | Clinell or 70% IPA wipe. The sleeve is single use. | Each patient |
| Docking station | Dry or lightly damp cloth, keeping the charging contacts dry | Weekly |
| USB cable | Dry cloth. Reconnect once the camera is fully dry. | Weekly |
Manual: GEIS user manual
Video Exam Glasses, Adjustable - USB-C
The clinician wears the glasses, and they don't normally touch the patient. Clean them between users and after any contact with the patient or their surroundings.
| Part | How to clean | How often |
|---|---|---|
| Camera module | Clinell or 70% IPA cloth, keeping liquid off the lens and out of the USB-C port | Between users |
| Lens | Dry lint-free cloth only, with minimal pressure | Before each use |
| Ball joint and clip | Clinell or 70% IPA cloth. Hold the joint while wiping so the angle doesn't move. | Between users |
| Frame or headband | Clinell or 70% IPA cloth. If clipped to your own glasses, clean those as usual. | Between users |
| Cable and connector | Dry or lightly damp cloth | Weekly |
Manuals: Examination glasses user manual | HD video camera glasses user guide
SyncVision Portable Digital Otoscope - USB
Semi-critical. Fit a new disposable speculum for every patient, and change it between ears if you suspect infection. Never examine an ear without a speculum fitted.
| Part | How to clean | How often |
|---|---|---|
| Disposable speculum | Single use. Remove and dispose of it after each patient. | Each patient |
| Tip and lens window | Check for debris. Dry lint-free cloth or lens tissue only, no liquid on the lens. | After each patient |
| Handle and body | Clinell or 70% IPA wipe, keeping liquid out of the USB port and buttons | After each patient |
| Charging cradle and cable | Dry cloth, keeping contacts dry | Weekly |
Use only the specula supplied for this otoscope. Other brands may not seat properly and can scratch the ear canal.
Manual: SyncVision i01 manual
CU2 Dental Intraoral Camera HD C-U2
Semi-critical. It goes in the mouth and touches mucous membranes and saliva, so a disposable sheath is mandatory for every patient.
| Part | How to clean | How often |
|---|---|---|
| Disposable sheath | Fit a new sheath before each patient. Wear gloves to remove it so the outside of the sheath doesn't touch the camera. Dispose of it as clinical waste. | Each patient |
| Handpiece and tip | After removing the sheath, wipe with the hospital-grade disinfectant wipe in the C-U2 IFU. Keep liquid off the lens window and out of joints. | After each patient |
| Lens window | Lens tissue or dry lint-free cloth | After each patient |
| Cable and USB connector | Dry or lightly damp cloth | Weekly |
| Holder | Clinell or 70% IPA wipe | Daily |
If a sheath tears during use, take the camera out of use until it's been cleaned and disinfected as the IFU sets out. Never immerse or autoclave the handpiece.
Manual: contact Visionflex support for the C-U2 IFU.
Related
- Cleaning and infection control for virtual care equipment
- Consumables and reordering
- Maintenance checklist, storage and repairs
Need help? Visit visionflex.com/support | email support@visionflex.com | call +61 2 8914 4000 (9am to 5pm AEST)
This is general guidance, not legal, clinical or financial advice. Always work within your scope of practice and your organisation's policies. Confirm your privacy and other legal obligations under the Privacy Act and current legislation, and seek professional advice for specific clinical, financial or medico-legal questions.